ICS Medical Advisory (ICSMA-20-296-02)

B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus

Click to Tweet.
Click to send to Facebook.
Click to Share.

Legal Notice

All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.


 

1. EXECUTIVE SUMMARY

  • CVSS v3 7.6
  • ATTENTION: Exploitable remotely/low and high skill level to exploit
  • Vendor: B. Braun Melsungen AG
  • Equipment: SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus
  • Vulnerabilities: Cross-site Scripting, Open Redirect, XPath Injection, Session Fixation, Use of a One-way Hash without a Salt, Relative Path Traversal, Improper Verification of Cryptographic Signature, Improper Privilege Management, Use of Hard-coded Credentials, Active Debug Code, Improper Access Control

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could allow an attacker to compromise the security of the Space or compactplus communication devices, allowing an attacker to escalate privileges, view sensitive information, upload arbitrary files, and perform remote code execution. 

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following versions of B. Braun products are affected:

  • SpaceCom, software Versions U61 and earlier (United States), L81 and earlier (outside the United States)
  • Battery pack with Wi-Fi, software Versions U61 and earlier (United States), L81 and earlier (outside the United States)
  • Data module compactplus, software Versions A10 and A11 (not distributed in the United States)

3.2 VULNERABILITY OVERVIEW

3.2.1    IMPROPER NEUTRALIZATION OF INPUT DURING WEB PAGE GENERATION ('CROSS-SITE SCRIPTING') CWE-79

A reflected cross-site scripting (XSS) vulnerability in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows remote attackers to inject arbitrary web script or HTML into various locations. 

CVE-2020-25158 has been assigned to this vulnerability. A CVSS v3 base score of 7.6 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:L/A:L).

3.2.2    URL REDIRECTION TO UNTRUSTED SITE ('OPEN REDIRECT') CWE-601

An open redirect vulnerability in the administrative interface of the B. Braun Melsungen AG SpaceCom device Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to redirect users to malicious websites. 

CVE-2020-25154 has been assigned to this vulnerability. A CVSS v3 base score of 5.4 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:L/A:N).

3.2.3    IMPROPER NEUTRALIZATION OF DATA WITHIN XPATH EXPRESSIONS ('XPATH INJECTION') CWE-643

A XPath injection vulnerability in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows unauthenticated remote attackers to access sensitive information and escalate privileges. 

CVE-2020-25162 has been assigned to this vulnerability. A CVSS v3 base score of 7.5 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N).

3.2.4    SESSION FIXATION CWE-384

A session fixation vulnerability in the B. Braun Melsungen AG SpaceCom administrative interface Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows remote attackers to hijack web sessions and escalate privileges. 

CVE-2020-25152 has been assigned to this vulnerability. A CVSS v3 base score of 6.5 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:N/A:N).

3.2.5    USE OF A ONE-WAY HASH WITHOUT A SALT CWE-759

A vulnerability in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to recover user credentials of the administrative interface. 

CVE-2020-25164 has been assigned to this vulnerability. A CVSS v3 base score of 5.1 has been assigned; the CVSS vector string is (AV:L/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N).

3.2.6    RELATIVE PATH TRAVERSAL CWE-23

A relative path traversal attack in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers with service user privileges to upload arbitrary files. By uploading a specially crafted tar file an attacker can execute arbitrary commands. 

CVE-2020-25150 has been assigned to this vulnerability. A CVSS v3 base score of 7.6 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:H/A:L).

3.2.7    IMPROPER VERIFICATION OF CRYPTOGRAPHIC SIGNATURE CWE-347

An improper verification of the cryptographic signature of firmware updates of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to generate valid firmware updates with arbitrary content that can be used to tamper with devices. 

CVE-2020-25166 has been assigned to this vulnerability. A CVSS v3 base score of 6.8 has been assigned; the CVSS vector string is (AV:P/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:H).

3.2.8    IMPROPER PRIVILEGE MANAGEMENT CWE-269

A vulnerability in the configuration import mechanism of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers with command line access to the underlying Linux system to escalate privileges to the root user. 

CVE-2020-16238 has been assigned to this vulnerability. A CVSS v3 base score of 6.7 has been assigned; the CVSS vector string is (AV:L/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H).

3.2.9    USE OF HARD-CODED CREDENTIALS CWE-798

Hard-coded credentials in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 enables attackers with command line access to access the device’s Wi-Fi module. 

CVE-2020-25168 has been assigned to this vulnerability. A CVSS v3 base score of 3.3 has been assigned; the CVSS vector string is (AV:L/AC:L/PR:L/UI:N/S:U/C:L/I:N/A:N).

3.2.10    ACTIVE DEBUG CODE CWE-489

Active debug code in the B. Braun Melsungen AG SpaceCom Version L8/U61, and the Data module compactplus Versions A10 and A11 and earlier enables attackers in possession of cryptographic material to access the device as root. 

CVE-2020-25156 has been assigned to this vulnerability. A CVSS v3 base score of 7.2 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H).

3.2.11    IMPROPER ACCESS CONTROL CWE-284

Improper access controls in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 enables attackers to extract and tamper with the devices network configuration. 

CVE-2020-25160 has been assigned to this vulnerability. A CVSS v3 base score of 6.8 has been assigned; the CVSS vector string is (AV:L/AC:L/PR:N/UI:N/S:C/C:L/I:L/A:L).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Germany

3.4 RESEARCHER

Julian Suleder, Nils Emmerich, and Birk Kauer of ERNW Research GmbH; Dr. Oliver Matula of ERNW Enno Rey Netzwerke GmbH, reported these vulnerabilities to the Federal Office for Information Security (BSI), Germany, in the context of the BSI project ManiMed (Manipulation of medical devices).

4. MITIGATIONS

B. Braun recommends applying updates:

  • SpaceCom: Version U62 or later (United States), L82 or later (outside the United States)
  • Battery Pack SP with Wi-Fi: Version U62 or later (United States), L82 or later (outside the United States)
  • Data module compactplus: Version A12 or later

As a general security measure, B. Braun recommends protecting the network with appropriate mechanisms:

  • Ensure the devices are not accessible directly from the Internet.
  • Use a firewall and isolate the medical devices from the business network. 

Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory.  

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability. Specifically, users should:

  • Minimize network exposure for all control system devices and/or systems, and ensure that they are not accessible from the Internet.
  • Locate control system networks and remote devices behind firewalls, and isolate them from the business network.
  • When remote access is required, use secure methods, such as Virtual Private Networks (VPNs), recognizing that VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize that VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on us-cert.cisa.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.cisa.gov in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.

No known public exploits specifically target these vulnerabilities. 


Contact Information

For any questions related to this report, please contact the CISA at:

Email: CISAservicedesk@cisa.dhs.gov
Toll Free: 1-888-282-0870

For industrial control systems cybersecurity information:  https://us-cert.cisa.gov/ics 
or incident reporting:  https://us-cert.cisa.gov/report

CISA continuously strives to improve its products and services. You can help by choosing one of the links below to provide feedback about this product.

This product is provided subject to this Notification and this Privacy & Use policy.

Please share your thoughts.

We recently updated our anonymous product survey; we'd welcome your feedback.