All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
1. EXECUTIVE SUMMARY
Biosense Webster Inc. (BWI), a Johnson & Johnson company, has produced a software update that applies operating system patches and anti-virus signature updates to close known vulnerabilities in the operating system of the CARTO 3 System, a 3D cardiovascular mapping platform. This update will be applied to CARTO 3 Systems beginning in April, 2018 as part of the free-of-charge CARTO 3 Version 6 (V6) base software version, which is designed to upgrade compatible CARTO 3 Systems running Version 4 (V4).
2. RISK EVALUATION
If the system is networked, the network interface for CARTO 3 V4 is sufficiently restricted by a software firewall to provide users reasonable assurance that it will not be exploited remotely or via malware/ransomware. If an attacker has persistent physical access to a CARTO 3 V4 System, the attacker could exploit the vulnerabilities in the operating system. This could allow the attacker to access information stored in the device, including individually identified health information about patients, affect the integrity of CARTO 3, or deny availability of the device. If the CARTO 3 V4 System is networked, an attacker with persistent physical access may also be able to access other systems within the user’s network.
Impact to individual organizations depends on many factors that are unique to each organization. NCCIC recommends that organizations evaluate the impact of these vulnerabilities based on their operational environment and specific clinical usage.
3. TECHNICAL DETAILS
3.1 AFFECTED PRODUCTS
The following CARTO 3 Systems versions are affected:
CARTO 3 Systems manufactured before April, 2018.
3.2 VULNERABILITY OVERVIEW
BWI reported controlled risks in the CARTO 3 System related to operating system vulnerabilities and outdated anti-virus signatures. A table providing CVE numbers, Microsoft vulnerability tracking numbers, and titles can be found at the following link:
http://www.productsecurity.jnj.com/CARTO3v4-Advisory-040918.pdf
3.2.1 VULNERABILITY DETAILS
3.2.2.1 EXPLOITABILITY
These vulnerabilities cannot be exploited remotely. Even if the CARTO 3 V4 System is networked, its network interface is restricted by a software firewall. These vulnerabilities require physical access to the CARTO 3 V4 System to exploit.
3.2.2.2 EXISTENCE OF EXPLOIT
Exploits that target these vulnerabilities exist and are publicly available.
3.2.2.3 DIFFICULTY
Exploiting these vulnerabilities would be difficult because an attacker must have physical access to the device and knowledge of the public exploits to exploit these vulnerabilities.
4.3 BACKGROUND
BWI, a Johnson & Johnson company, is a U.S.-based company that maintains offices in several countries around the world, including the U.S., Asia, Europe, Middle East, and Africa.
The affected product, the CARTO 3 V4 System, is an imaging device that uses electromagnetic technology to create real-time three-dimensional (3D) maps of a patient’s cardiac structures. According to BWI, CARTO 3 is deployed across the healthcare and public health sector. BWI estimates these products are used primarily in the United States, Asia, Europe, Middle East, and Africa.
5. MITIGATIONS
Physical security for the CARTO 3 System is a critical control that must be employed by device users to limit the exposure of identified risks and vulnerabilities that can be exploited with persistent physical access to the device.
BWI will be contacting users to initiate a software update in the field to address vulnerabilities within the CARTO 3 System.
Users with questions regarding the CARTO 3 system are advised to contact their BWI sales representative or service technician.
Please see the Johnson & Johnson Product Security website for the latest security information for Products of the Johnson & Johnson Family of Companies:
http://productsecurity.jnj.com
NCCIC recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
- Minimize network exposure for all control system devices and/or systems, and ensure that they are not accessible from the Internet.
- Locate all medical devices and remote devices behind firewalls, and isolate them from the business network.
- When remote access is required, use secure methods, such as Virtual Private Networks (VPNs), recognizing that VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize that VPN is only as secure as the connected devices.
NCCIC also provides a section for control systems security recommended practices on the ICS-CERT web page. NCCIC reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
Additional mitigation guidance and recommended practices are publicly available in the NCCIC Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies, that is available for download from the ICS-CERT website.
Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to NCCIC for tracking and correlation against other incidents.
Contact Information
For any questions related to this report, please contact the CISA at:
Email: CISAservicedesk@cisa.dhs.gov
Toll Free: 1-888-282-0870
For industrial control systems cybersecurity information: https://us-cert.cisa.gov/ics
or incident reporting: https://us-cert.cisa.gov/report
CISA continuously strives to improve its products and services. You can help by choosing one of the links below to provide feedback about this product.
This product is provided subject to this Notification and this Privacy & Use policy.
Please share your thoughts.
We recently updated our anonymous product survey; we'd welcome your feedback.