Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
OVERVIEW
Philips reported vulnerabilities in the Philips’ IntelliSpace Portal (ISP), an advanced visualization and image analysis system. Philips is creating a software update to mitigate these vulnerabilities in the affected products. Additionally, they are issuing mitigating controls for some vulnerabilities.
Some vulnerabilities could be exploited remotely.
Exploits that target some vulnerabilities are publicly available.
AFFECTED PRODUCTS
Philips reports that these vulnerabilities affect the following versions of the ISP:
- IntelliSpace Portal, all 8.0.x versions, and
- IntelliSpace Portal, all 7.0.x versions.
IMPACT
Successful exploitation of these vulnerabilities could allow an attacker to gain unauthorized access to sensitive information, perform man-in-the-middle attacks, create denial of service conditions, or execute arbitrary code.
Impact to individual organizations depends on many factors that are unique to each organization. NCCIC recommends that organizations evaluate the impact of these vulnerabilities based on their operational environment and specific clinical usage.
BACKGROUND
Philips is a global company that maintains offices in many countries around the world, including countries in Africa, Asia, Europe, Latin America, the Middle East, and North America.
The Philips ISP processes clinical images from different modalities and enables advanced visualization of the images. ISP systems are deployed across the Healthcare and Public Health sectors. Philips estimates these products are used worldwide.
VULNERABILITY CHARACTERIZATION
VULNERABILITY OVERVIEW
The ISP has multiple input validation vulnerabilities that could allow a remote attacker to execute arbitrary code or cause the application to crash.
CVE-2018-5474, CVE-2017-0143, CVE-2017-0144, CVE-2017-0145, CVE-2017-0146, CVE-2017-0148, CVE-2017-0272, CVE-2017-0277, CVE-2017-0278, CVE-2017-0279, CVE-2017-0269, CVE-2017-0273, and CVE-2017-0280 have been assigned to these vulnerabilities. The CVSS v3 base scores for theses vulnerabilities range from 5.9 to 8.1
The ISP has multiple information exposure vulnerabilities that could allow an attacker to gain unauthorized access to sensitive information.
CVE-2017-0147, CVE-2017-0267, CVE-2017-0268, CVE-2017-0270, CVE-2017-0271, CVE-2017-0274, CVE-2017-0275, and CVE-2017-0276, have been assigned to these vulnerabilities. A CVSS v3 base score of 5.9 has been calculated; the CVSS vector string is (AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N).
The ISP has multiple permission, privilege and access control vulnerabilities that could allow an attacker to gain unauthorized access and in some cases escalate their level of privilege or execute arbitrary code.
CVE-2018-5472, CVE-2018-5468, CVE-2017-0199, and CVE-2005-1794 have been assigned to this vulnerability. The CVSS v3 base scores for theses vulnerabilities range from 6.4 to 7.8
An unquoted search path or element vulnerability has been identified, which may allow an authorized local user to execute arbitrary code and escalate their level of privileges.
CVE-2018-5470, has been assigned to this vulnerability. A CVSS v3 base score of 7.8 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H).
The ISP has a vulnerability where code debugging methods are enabled, which could allow an attacker to remotely execute arbitrary code during runtime.
CVE-2018-5454 has been assigned to this vulnerability. A CVSS v3 base score of 5.3 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:L/A:N).
The ISP has multiple cryptographic vulnerabilities that could allow an attacker to gain unauthorized access to resources and information.
CVE-2018-5458, CVE-2018-5462, CVE-2018-5464, CVE-2018-5466, CVE-2011-3389, CVE-2004-2761, CVE-2014-3566, and CVE-2016-2183 have been assigned to these vulnerabilities. The CVSS v3 base scores for theses vulnerabilities range from 3.1 to 6.5
VULNERABILITY DETAILS
EXPLOITABILITY
Some vulnerabilities could be exploited remotely.
EXISTENCE OF EXPLOIT
Public exploits exist for some of these vulnerabilities; however, none are known to specifically target Philips ISP.
DIFFICULTY
An attacker with a low skill would be able to exploit these vulnerabilities.
MITIGATION
Philips will release an updated version of the ISP in the coming months that will address these vulnerabilities. Additionally, Philips’ evaluation of Operating System security patches is ongoing, and after appropriate testing , the patches and mitigating controls are posted on Philips’ InCenter. ISP users are recommended to obtain available mitigating controls by accessing their InCenter account at this location:
http://incenter.medical.philips.com
Users with questions regarding their specific ISP installations are advised by Philips to contact their local Philips service support team or their regional service support.
Philips’ contact information is available at the following location:
https://www.usa.philips.com/healthcare/solutions/customer-service-solutions
Please see the Philips product security website for the latest security information for Philips products:
https://www.philips.com/productsecurity
NCCIC recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
- Minimize network exposure for all control system devices and/or systems, and ensure that they are not accessible from the Internet.
- Locate all medical devices and remote devices behind firewalls, and isolate them from the business network.
- When remote access is required, use secure methods, such as Virtual Private Networks (VPNs), recognizing that VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize that VPN is only as secure as the connected devices.
NCCIC also provides a section for control systems security recommended practices on the ICS-CERT web page. NCCIC reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
Additional mitigation guidance and recommended practices are publicly available in the NCCIC Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies, that is available for download from the ICS-CERT website.
Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to NCCIC for tracking and correlation against other incidents.
Contact Information
For any questions related to this report, please contact the CISA at:
Email: CISAservicedesk@cisa.dhs.gov
Toll Free: 1-888-282-0870
For industrial control systems cybersecurity information: https://us-cert.cisa.gov/ics
or incident reporting: https://us-cert.cisa.gov/report
CISA continuously strives to improve its products and services. You can help by choosing one of the links below to provide feedback about this product.
This product is provided subject to this Notification and this Privacy & Use policy.
Please share your thoughts.
We recently updated our anonymous product survey; we'd welcome your feedback.