Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
OVERVIEW
Siemens has identified four vulnerabilities in Siemens’ Molecular Imaging products running on Windows 7. Siemens is preparing updates for the affected products.
These vulnerabilities could be exploited remotely. Exploits that target these vulnerabilities are known to be publicly available.
AFFECTED PRODUCTS
Siemens reports that the vulnerabilities affect the following products:
- Siemens PET/CT Systems: All Windows 7-based versions,
- Siemens SPECT/CT Systems: All Windows 7-based versions,
- Siemens SPECT Systems: All Windows 7-based versions, and
- Siemens SPECT Workplaces/Symbia.net: All Windows 7-based versions.
IMPACT
Successful exploitation of these vulnerabilities may allow the attacker to remotely execute arbitrary code.
Impact to individual organizations depends on many factors that are unique to each organization. NCCIC/ICS-CERT recommends that organizations evaluate the impact of these vulnerabilities based on their operational environment and specific clinical usage.
BACKGROUND
Siemens is an international company headquartered in Munich, Germany.
The affected products, Select Molecular Imaging products, are used in medical imaging. According to Siemens, Molecular Imaging products are deployed across the Healthcare and Public Health sector. Siemens estimates that these products are used worldwide.
VULNERABILITY CHARACTERIZATION
VULNERABILITY OVERVIEW
IMPROPER CONTROL OF GENERATION OF CODE ('CODE INJECTION')a
An unauthenticated remote attacker could execute arbitrary code by sending specially crafted HTTP requests to the Microsoft web server (Port 80/TCP and Port 443/TCP) of affected devices.
CVE-2015-1635b has been assigned to this vulnerability. A CVSS v3 base score of 9.8 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).c
IMPROPER CONTROL OF GENERATION OF CODE ('CODE INJECTION')d
An unauthenticated remote attacker could execute arbitrary code by sending a specially crafted request to the HP Client automation service on Port 3465/TCP of affected devices.
CVE-2015-1497e has been assigned to this vulnerability. A CVSS v3 base score of 9.8 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).f
IMPROPER RESTRICTION OF OPERATIONS WITHIN THE BOUNDS OF A MEMORY BUFFERg
An unauthenticated remote attacker could execute arbitrary code by sending a specially crafted request to the HP Client automation service of affected devices.
CVE-2015-7860h has been assigned to this vulnerability. A CVSS v3 base score of 9.8 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).i
PERMISSIONS, PRIVILEGES, AND ACCESS CONTROLSj
An unauthenticated remote attacker could execute arbitrary code by sending a specially crafted request to the HP Client automation service of affected devices.
CVE-2015-7861k has been assigned to this vulnerability. A CVSS v3 base score of 9.8 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).l
VULNERABILITY DETAILS
EXPLOITABILITY
These vulnerabilities could be exploited remotely.
EXISTENCE OF EXPLOIT
Exploits that target these vulnerabilities are publicly available.
DIFFICULTY
An attacker with a low skill would be able to exploit these vulnerabilities.
MITIGATION
Siemens is preparing updates for the affected products and recommends protecting network access to the Molecular Imaging products with appropriate mechanisms. It is advised to run the devices in a dedicated network segment and protected IT environment. If this is not possible, Siemens recommends the following:
- If patient safety and treatment is not at risk, disconnect the product from the network and use in standalone mode.
- Reconnect the product only after the provided patch or remediation is installed on the system. Siemens is able to patch systems capable of Remote Update Handling (RUH) much faster by remote software distribution compared to onsite visits. Therefore users of RUH-capable equipment are recommended to clarify the situation concerning patch availability and remaining risk in the local customer network with the Siemens Customer Care Center first and then to reconnect the systems in order to receive patches as quickly as possible via RUH. This ensures smooth and fast receipt of updates and therefore supports reestablishment of system operations.
In addition, Siemens recommends:
- Ensure users have appropriate backups and system restoration procedures.
- For specific patch and remediation guidance information contact a local Siemens customer service engineer or a Siemens regional support center.
For more information on these vulnerabilities and more detailed mitigation instructions, please see Siemens Security Advisory SSA-822184 at the following location:
http://www.siemens.com/cert/advisories
ICS-CERT recommends that users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
- Minimize network exposure for all medical devices and/or systems, and ensure that they are not accessible from the Internet.
- Locate all medical devices and remote devices behind firewalls, and isolate them from the business network.
- When remote access is required, use secure methods, such as Virtual Private Networks (VPNs), recognizing that VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize that VPN is only as secure as the connected devices.
ICS-CERT also provides a section for security recommended practices on the ICS-CERT web page at http://ics-cert.us-cert.gov/content/recommended-practices. ICS-CERT reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
Additional mitigation guidance and recommended practices are publicly available in the ICS‑CERT Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies, that is available for download from the ICS-CERT web site (http://ics-cert.us-cert.gov/).
Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to ICS-CERT for tracking and correlation against other incidents.
- aCWE-94: Improper Control of Generation of Code ('Code Injection'), https://cwe.mitre.org/data/definitions/94.html, web site last accessed August 03, 2017.
- bNVD, https://nvd.nist.gov/vuln/detail/CVE-2015-1635, NIST uses this advisory to create the CVE web site report. This web site will be active sometime after publication of this advisory.
- cCVSS Calculator, https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H, web site last accessed August 03, 2017.
- dCWE-94: Improper Control of Generation of Code ('Code Injection'), https://cwe.mitre.org/data/definitions/94.html, web site last accessed August 03, 2017.
- eNVD, https://nvd.nist.gov/vuln/detail/CVE-2015-1497, NIST uses this advisory to create the CVE web site report. This web site will be active sometime after publication of this advisory.
- fCVSS Calculator, https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H, web site last accessed August 03, 2017.
- gCWE-119: Improper Restriction of Operations within the Bounds of a Memory Buffer, https://cwe.mitre.org/data/definitions/119.html, web site last accessed August 03, 2017.
- hNVD, https://nvd.nist.gov/vuln/detail/CVE-2015-7860, NIST uses this advisory to create the CVE web site report. This web site will be active sometime after publication of this advisory.
- iCVSS Calculator, https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H, web site last accessed August 03, 2017.
- jCWE-264: Permissions, Privileges, and Access Controls, https://cwe.mitre.org/data/definitions/264.html, web site last accessed August 03, 2017.
- kNVD, https://nvd.nist.gov/vuln/detail/CVE-2015-7861, NIST uses this advisory to create the CVE web site report. This web site will be active sometime after publication of this advisory.
- lCVSS Calculator, https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H, web site last accessed August 03, 2017.
Contact Information
For any questions related to this report, please contact the CISA at:
Email: CISAservicedesk@cisa.dhs.gov
Toll Free: 1-888-282-0870
For industrial control systems cybersecurity information: https://us-cert.cisa.gov/ics
or incident reporting: https://us-cert.cisa.gov/report
CISA continuously strives to improve its products and services. You can help by choosing one of the links below to provide feedback about this product.
This product is provided subject to this Notification and this Privacy & Use policy.
Please share your thoughts.
We recently updated our anonymous product survey; we'd welcome your feedback.