Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
OVERVIEW
This updated advisory is a follow-up to the original advisory titled ICSMA-17-009-01 St. Jude Merlin@home Transmitter Vulnerability that was published January 9, 2017, on the NCCIC/ICS-CERT web site.
--------- Begin Update A Part 1 of 5 --------
MedSec Holdings has identified a channel accessible by nonendpoint (“man-in-the-middle”) vulnerability in St. Jude Medical's Merlin@home transmitter, which affects both the RF and inductive models. St. Jude Medical has validated the vulnerability and produced a new software version that mitigates this vulnerability. A third-party security research firm has verified that the new software version mitigates the identified vulnerability.
--------- End Update A Part 1 of 5 ----------
This vulnerability could be exploited remotely. An attacker with high skill would be able to exploit this vulnerability.
The Food and Drug Administration (FDA) released a safety communication on January 9, 2017, Cybersecurity Vulnerabilities Identified in St. Jude Medical’s Implantable Cardiac Devices and Merlin@home Transmitter, to alert users about the identified vulnerability and corresponding mitigation as well as to provide recommendations to patients and healthcare providers. In response, ICS-CERT is releasing this advisory to provide additional information to patients and healthcare providers.
AFFECTED PRODUCTS
--------- Begin Update A Part 2 of 5 --------
The following Merlin@home transmitters are affected:
- Merlin@home, versions prior to Version 8.2.2:
- RF models: EX1150,
- Inductive models: EX1100, and
- Inductive models: EX1100 with MerlinOnDemand capability.
--------- End Update A Part 2 of 5 ----------
IMPACT
Successful exploitation of this vulnerability may allow a remote attacker to access or influence communications between Merlin.net and transmitter endpoints.
ICS-CERT recommends that patients and healthcare providers evaluate the impact of this vulnerability based on their specific usage after reviewing the information referenced in this advisory and to contact the vendor for assistance with any questions or concerns related to this vulnerability.
BACKGROUND
St. Jude Medical is a US-based company headquartered in St. Paul, Minnesota.
The affected product, the Merlin@home transmitter, allows for remote care management of patients with implanted cardiac devices through scheduled transmissions, patient-initiated transmissions, and daily monitoring.
--------- Begin Update A Part 3 of 5 --------
According to St. Jude Medical, Merlin@home transmitter inductive models can have the MerlinOnDemand capability enabled. The MerlinOnDemand capability is a configuration that allows for multiple-patient use in healthcare facilities. The Merlin@home transmitter inductive models, with MerlinOnDemand capability enabled, is used by healthcare professionals to read patient Implantable Medical Devices (IMDs) data during office visits. Inductive Merlin@home transmitters with MerlinOnDemand capability enabled contain the same hardware and software as other inductive Merlin@home transmitters, but with slightly different capability, resulting from a different configuration on the server/Merlin.net side, which is controlled by St. Jude Medical. The Merlin@home transmitter inductive model with the MerlinOnDemand capability enabled is set up to pair with multiple IMDs, which differs from the Merlin@home transmitter that has not enabled this capability, as it is intended for home use. Also, the Merlin@home transmitter inductive model with MerlinOnDemand capability enabled does not use RF to communicate with IMDs. Instead, it uses close range impedance telemetry.
According to St. Jude Medical, the Merlin@home transmitter inductive models are deployed across the Healthcare and Public Health sector. St. Jude Medical estimates that Merlin@home transmitters inductive models, with MerlinOnDemand capability enabled, correspond to approximately 0.1 percent of the total Merlin@home transmitters available and that these products are used worldwide.
--------- End Update A Part 3 of 5 ----------
VULNERABILITY CHARACTERIZATION
VULNERABILITY OVERVIEW
CHANNEL ACCESSIBLE BY NONENDPOINT (“MAN-IN-THE-MIDDLE”)a
--------- Begin Update A Part 4 of 5 --------
The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical’s web site, Merlin.net, are not verified. This may allow a remote attacker to access or influence communications between the identified endpoints.
--------- End Update A Part 4 of 5 ----------
CVE-2017-5149b has been assigned to this vulnerability. A CVSS v3 base score of 8.9 has been assigned; the CVSS vector string is (AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:H).c
VULNERABILITY DETAILS
EXPLOITABILITY
This vulnerability could be exploited remotely.
EXISTENCE OF EXPLOIT
No known public exploits specifically target this vulnerability.
DIFFICULTY
An attacker with high skill would be able to exploit this vulnerability.
MITIGATION
--------- Begin Update A Part 5 of 5 --------
St. Jude Medical has developed an updated software version for all Merlin@home transmitters, including inductive Merlin@home transmitters with the MerlinOnDemand capability, which mitigates the identified vulnerability and provides additional security enhancements.
The new version of the transmitter software, Version 8.2.2, will be automatically updated over a period of several months, when all models of the Merlin@home transmitters are connected to an Ethernet, WiFi, cellular network, or a landline. St. Jude Medical recommends that users keep Merlin@home transmitters powered and connected at all times to receive this update and future updates.
For additional information about the vulnerability or the software update process, users can review information from St. Jude Medical at:
Patients and healthcare providers with questions can call the Merlin hotline at
1-877-696-3754 or visit www.sjm.com/Merlin for more information.
The FDA issued a safety communication on January 9, 2017, Cybersecurity Vulnerabilities Identified in St. Jude Medical’s Implantable Cardiac Devices and Merlin@home Transmitter, which includes recommendations for patients and healthcare providers and is available at the following location:
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm535843.htm
--------- End Update A Part 5 of 5 ----------
St. Jude Medical is continuing to work with ICS-CERT and the FDA to address additional security issues that have been identified. As additional information becomes available, ICS-CERT in coordination with the FDA, will release additional information products.
ICS-CERT reminds Internet users that directly connecting any device to the Internet without explicitly controlling communication with or access to the connected device, significantly increases the risk of a cybersecurity-related event.
ICS-CERT provides recommended security practices on the ICS-CERT web page at http://ics-cert.us-cert.gov/content/recommended-practices. ICS-CERT reminds members of the healthcare and public health sector to perform proper impact analysis and risk assessment prior to deploying defensive measures.
Additional mitigation guidance and recommended practices are publicly available in the ICS‑CERT Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies, that is available for download from the ICS-CERT web site (http://ics-cert.us-cert.gov/).
- aCWE-300: Channel Accessible by Non-Endpoint ('Man-in-the-Middle'), http://cwe.mitre.org/data/definitions/300.html, web site last accessed January 9, 2017.
- bNVD, http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2017-5149, NIST uses this advisory to create the CVE web site report. This web site will be active sometime after publication of this advisory.
- cCVSS Calculator, https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S..., web site last accessed January 9, 2017.
Contact Information
For any questions related to this report, please contact the CISA at:
Email: CISAservicedesk@cisa.dhs.gov
Toll Free: 1-888-282-0870
For industrial control systems cybersecurity information: https://us-cert.cisa.gov/ics
or incident reporting: https://us-cert.cisa.gov/report
CISA continuously strives to improve its products and services. You can help by choosing one of the links below to provide feedback about this product.
This product is provided subject to this Notification and this Privacy & Use policy.
Please share your thoughts.
We recently updated our anonymous product survey; we'd welcome your feedback.