Legal Notice
All information products included in https://us-cert.cisa.gov/ics are provided "as is" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.
OVERVIEW
This advisory was originally posted to the US-CERT secure Portal library on June 30, 2015, and is being released to the NCCIC/ICS-CERT web site.
Researcher Jared Bird with Allina IS Security identified four vulnerabilities in Baxter’s SIGMA Spectrum Infusion System. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8 that incorporates hardware and software changes, which do not contain three of the four identified vulnerabilities.
Three of the four vulnerabilities are remotely exploitable.
AFFECTED PRODUCTS
The following SIGMA Spectrum Infusion System versions are affected:
- SIGMA Spectrum Infusion System, Version 6.05 (model 35700BAX) with wireless battery module (WBM), Version 16. The WBM is a stand-alone component that provides network connectivity to the pump.
IMPACT
Successful exploitation of these vulnerabilities may allow a remote attacker to make unauthorized configuration changes to the WBM and gain information about the host network such as wireless account credentials. According to Baxter, it is not possible to change infusion parameters using the identified vulnerabilities. In addition, the SIGMA Spectrum Infusion Pump does not contain any personally identifiable information or patient health information.
Impact to individual organizations depends on many factors that are unique to each organization. ICS-CERT recommends that organizations evaluate the impact of these vulnerabilities based on their operational environment and specific clinical usage.
BACKGROUND
Baxter is a US-based company that maintains offices worldwide, including the US, UK, Italy, India, Germany, France, China, and Australia.
The affected product, the SIGMA Spectrum Infusion System, is an intravenous pump that delivers medication to patients. According to Baxter, SIGMA Spectrum Infusion Systems are deployed across the Healthcare and Public Health sector. Baxter estimates that these products are used in the US and Canada.
VULNERABILITY CHARACTERIZATION
VULNERABILITY OVERVIEW
USE OF HARD-CODED PASSWORDa
Baxter’s SIGMA Spectrum infusion pumps contain a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase.
CVE-2014-5431b has been assigned to this vulnerability. A CVSS v2 base score of 4.6 has been assigned; the CVSS vector string is (AV:L/AC:L/Au:N/C:P/I:P/A:P).c
AUTHENTICATION BYPASS ISSUESd
The WBM is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM.
CVE-2014-5432e has been assigned to this vulnerability. A CVSS v2 base score of 7.5 has been assigned; the CVSS vector string is (AV:N/AC:L/Au:N/C:P/I:P/A:P).f
CLEARTEXT STORAGE OF SENSITIVE INFORMATIONg
An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on the WBM, which may allow an attacker to gain access the host network.
CVE-2014-5433h has been assigned to this vulnerability. A CVSS v2 base score of 9.0 has been assigned; the CVSS vector string is (AV:N/AC:L/Au:N/C:C/I:P/A:P).i
USE OF HARD-CODED PASSWORDj
The WBM has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account.
CVE-2014-5434k has been assigned to this vulnerability. Baxter has assigned a CVSS v2 base score of 5.0; the CVSS vector string is (AV:N/AC:L/Au:N/C:P/I:N/A:N).l
VULNERABILITY DETAILS
EXPLOITABILITY
Three of the four vulnerabilities could be exploited remotely. Exploitation of the hard-coded password vulnerability requires local access.
EXISTENCE OF EXPLOIT
No known public exploits specifically target these vulnerabilities.
DIFFICULTY
An attacker with a low skill level would be able to exploit these vulnerabilities.
MITIGATION
Baxter offers the following recommendations to help mitigate risks associated with these vulnerabilities in the SIGMA Spectrum Infusion System running Version 6.05 with WBM Version 16.
- Ensure that the WI-FI network supporting WBMs is secured using a secure WI-FI protocol.
- Separate the network supporting the WBMs with a standalone VLAN or use similarly segmented network topography to isolate WBMs. This would require an attacker to compromise the standalone WI-FI network or otherwise gain access to the supporting VLAN before SSH access to the WBM is possible.
- Configure Wireless Access Points and Firewalls, which provide access to the VLAN, to block Port 21/FTP and Port 22/SSH.
- Ensure that network authentication credentials used by the WBM to connect to the network are properly restricted to only allow access to the wireless network.
- As a last resort, customers may disable wireless operation of the pump. The Sigma Spectrum Infusion System was designed to operate without network access. This action would impact an organization’s ability to rapidly deploy drug library (formulary) updates to their pumps.
Baxter states that it has implemented a process to continually evaluate cybersecurity risks and has defined a roadmap to mitigate vulnerabilities. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes that do not contain three of the four identified vulnerabilities. In Version 8, Baxter has addressed the authentication bypass issue by removing the SSH service from the WBM. The new version addresses the clear text storage of sensitive information through modifications to the commands used to expose network and WI-FI credentials on the WBM; security key information is now masked or otherwise removed from command outputs. Furthermore, the path to gain access to these commands is closed, as the SSH service has been removed. In Version 8, Baxter has addressed the FTP hard-coded password vulnerability by removing the FTP service from the WBM. Baxter has engaged an independent security expert to confirm that Version 8 does not contain the three remotely exploitable vulnerabilities.
Baxter has performed a cybersecurity risk analysis and has evaluated the potential impact of the hard-coded password to access the device as being low. Baxter plans to address this in a future release. Baxter recommends that facilities employ physical security controls to ensure the safety of the pump and WBM.
For additional information about the vulnerabilities, compensating measures, or the new version of the SIGMA Spectrum Infusion System, contact Baxter Technical Support at: 1-800-843-7867 or via email at: gts@baxter.com.
ICS-CERT encourages asset owners to take additional defensive measures to protect against this and other cybersecurity risks.
- Minimize network exposure for all medical devices and/or systems and ensure that they are not accessible from the Internet.
- Locate all medical devices and/or systems behind firewalls and isolate them from the business network.
- When remote access is required, use secure methods, such as Virtual Private Networks (VPNs), recognizing that VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize that VPN is only as secure as the connected devices.
ICS-CERT also provides a section for security recommended practices on the ICS-CERT web page at: http://ics-cert.us-cert.gov/content/recommended-practices. ICS-CERT reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
Additional mitigation guidance and recommended practices are publicly available in the ICS-CERT Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies, that is available for download from the ICS-CERT web site (http://ics-cert.us-cert.gov/).
Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to ICS-CERT for tracking and correlation against other incidents.
- a. CWE-259: Use of Hard-coded Password, http://cwe.mitre.org/data/definitions/259.html, web site last accessed June 30, 2015.
- b. NVD, http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2014-5431, NIST uses this advisory to create the CVE web site report. This web site will be active sometime after publication of this advisory.
- c. CVSS Calculator, http://nvd.nist.gov/cvss.cfm?version=2&vector=AV:L/AC:L/Au:N/C:P/I:P/A:P, web site last accessed June 30, 2015.
- d. CWE-592: Authentication Bypass Issues, http://cwe.mitre.org/data/definitions/592.html, web site last accessed June 30, 2015.
- e. NVD, http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2014-5432, NIST uses this advisory to create the CVE web site report. This web site will be active sometime after publication of this advisory.
- f. CVSS Calculator, http://nvd.nist.gov/cvss.cfm?version=2&vector=AV:N/AC:L/Au:N/C:P/I:P/A:P, web site last accessed June 30, 2015.
- g. CWE-312: Cleartext Storage of Sensitive Information, http://cwe.mitre.org/data/definitions/312.html, web site last accessed June 30, 2015.
- h. NVD, http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2014-5433, NIST uses this advisory to create the CVE web site report. This web site will be active sometime after publication of this advisory.
- i. CVSS Calculator, http://nvd.nist.gov/cvss.cfm?version=2&vector=AV:N/AC:L/Au:N/C:C/I:P/A:P, web site last accessed June 30, 2015.
- j. CWE-259: Use of Hard-coded Password, http://cwe.mitre.org/data/definitions/259.html, web site last accessed June 30, 2015.
- k. NVD, http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2014-5434, NIST uses this advisory to create the CVE web site report. This web site will be active sometime after publication of this advisory.
- l. CVSS Calculator, http://nvd.nist.gov/cvss.cfm?version=2&vector=AV:N/AC:L/Au:N/C:P/I:N/A:N, web site last accessed June 30, 2015.
Contact Information
For any questions related to this report, please contact the CISA at:
Email: CISAservicedesk@cisa.dhs.gov
Toll Free: 1-888-282-0870
For industrial control systems cybersecurity information: https://us-cert.cisa.gov/ics
or incident reporting: https://us-cert.cisa.gov/report
CISA continuously strives to improve its products and services. You can help by choosing one of the links below to provide feedback about this product.
This product is provided subject to this Notification and this Privacy & Use policy.
Please share your thoughts.
We recently updated our anonymous product survey; we'd welcome your feedback.